PR9112 Calibrator Pressurae Intelligens
Parametri Technici
| Modellum | PR9112Calibrator Pressionis Intelligens | |
| Mensura Pressionis | Spatium Mensurae | (-0.1~250)Mpa |
| Accuratio ostentationis | ±0.05% FS, ±0.02% FS | |
| Mensura Currentis Electrici | Spatium | ±30.0000mA |
| Sensibilitatem | 0.1µA | |
| Accuratio | ± (0.01% R.D + 0.003% FS) | |
| Mensura Tensionis | Spatium | ±30.0000V |
| Sensibilitatem | 0.1mV | |
| Accuratio | ± (0.01% RD + 0.003% FS) | |
| Valor Commutans | Grex mensurarum potentiae/interruptionis | |
| Functio emissionis | Egressus currentis continui | DC24V±0.5V |
| Operandi Ambitus | Temperatura Operativa | (-20~50)℃ |
| Temperatura Relativa | <95% | |
| Temperatura Reponendi | (-30~80)℃ | |
| Configuratio Fontis Electrici | Modus alimentationis electricae | Accumulator lithii vel fons potentiae |
| Tempus Operationis Batteriae | Horae LX (24V sine onere) | |
| Tempus impletionis | Circiter quattuor horas | |
| Alia indicia | Magnitudo | 115mm × 45mm × 180mm |
| Interfacies communicationis | obturaculum aviationis trium filorum specializatum | |
| Pondus | 0.8 kg | |
Applicatio Praecipua:
1. Transmissorem pressionis (pressionis differentialis) calibrare
2. Pressostat calibrare
3. Manometrum accuratum, manometrum generalem, verifica.
Proprietas Producti:
1. Functio operatoris manualis inclusa, transmissor pressionis intelligens HART calibrari potest (ad libitum).
2. Ostentus crystalli liquidi duplicis ordinis cum illuminatione posteriori.
3. mmH2O, mmHg, psi, kPa, MPa, Pa, mbar, bar, kgf/c㎡, inter novem unitates pressionis commuta.
4. Cum functione egressus DC24V.
5. Cum mensura currentis et tensionis.
6. Mensura cum volumine commutante.
7. Cum interfacie communicationis (ad libitum)
8. Capacitas repositionis: summa fasciculi 30, (singulorum fasciculorum 50 inscriptiones datorum)
9. Magnum velum crystallinum liquidum ostentat
Configuratio Programmatum:
Programma PR9112S systematis verificationis pressionis est programmatum adiuvans systematis nostri digitalis.calibrator pressionisSeries productorum in societate nostra, acta collectionis datorum fieri possunt, forma automatice generata, computatio errorum automatica, impressio certificati.
1.Tabula Selectionis Intervalli Pressionis Ordinariae
| Non. | Ambitus Pressionis | Typus | Classis accurationis |
| 01 | (-100~0) kPa | G | 0.02/0.05 |
| 02 | (0~60)Pa | G | 0.2/0.05 |
| 03 | (0~250)Pa | G | 0.2/0.05 |
| 04 | (0 ~ 1) kPa | G | 0.05/0.1 |
| 05 | (0 ~ 2) kPa | G | 0.05/0.1 |
| 06 | (0 ~ 2.5) kPa | G | 0.05/0.1 |
| 07 | (0 ~ 5) kPa | G | 0.05/0.1 |
| 08 | (0 ~ 10) kPa | G | 0.05/0.1 |
| 09 | (0 ~ 16) kPa | G | 0.05/0.1 |
| 10 | (0 ~ 25) kPa | G | 0.05/0.1 |
| 11 | (0 ~ 40) kPa | G | 0.05/0.1 |
| 12 | (0 ~ 60) kPa | G | 0.05/0.1 |
| 13 | (0 ~ 100) kPa | G | 0.05/0.1 |
| 14 | (0 ~ 160) kPa | Generalis Litterarum | 0.02/0.05 |
| 15 | (0 ~ 250) kPa | Generalis Litterarum | 0.02/0.05 |
| 16 | (0 ~ 400) kPa | Generalis Litterarum | 0.02/0.05 |
| 17 | (0 ~ 600) kPa | Generalis Litterarum | 0.02/0.05 |
| 18 | (0 ~ 1) MPa | Generalis Litterarum | 0.02/0.05 |
| 19 | (0 ~ 1.6) MPa | Generalis Litterarum | 0.02/0.05 |
| 20 | (0 ~ 2.5) MPa | Generalis Litterarum | 0.02/0.05 |
| 21 | (0 ~ 4) MPa | Generalis Litterarum | 0.02/0.05 |
| 22 | (0 ~ 6) MPa | Generalis Litterarum | 0.02/0.05 |
| 23 | (0 ~ 10) MPa | Generalis Litterarum | 0.02/0.05 |
| 24 | (0 ~ 16) MPa | Generalis Litterarum | 0.02/0.05 |
| 25 | (0 ~ 25) MPa | Generalis Litterarum | 0.02/0.05 |
| 26 | (0 ~ 40) MPa | Generalis Litterarum | 0.02/0.05 |
| 27 | (0 ~ 60) MPa | Generalis Litterarum | 0.05/0.1 |
| 28 | (0 ~ 100) MPa | Generalis Litterarum | 0.05/0.1 |
| 29 | (0 ~ 160) MPa | Generalis Litterarum | 0.05/0.1 |
| 30 | (0 ~ 250) MPa | Generalis Litterarum | 0.05/0.1 |
Adnotationes: G = Gas L = Liquidum
2.Tabula Selectionis Intervalli Pressionis Compositae:
| Non. | Ambitus Pressionis | Typus | Classis accurationis |
| 01 | ±60 Pa | G | 0.2/0.5 |
| 02 | ±160 Pa | G | 0.2/0.5 |
| 03 | ±250 Pa | G | 0.2/0.5 |
| 04 | ±500 Pa | G | 0.2/0.5 |
| 05 | ±1kPa | G | 0.05/0.1 |
| 06 | ±2kPa | G | 0.05/0.1 |
| 07 | ±2.5 kPa | G | 0.05/0.1 |
| 08 | ±5kPa | G | 0.05/0.1 |
| 09 | ±10kPa | G | 0.05/0.1 |
| 10 | ±16kPa | G | 0.05/0.1 |
| 11 | ±25kPa | G | 0.05/0.1 |
| 12 | ±40kPa | G | 0.05/0.1 |
| 13 | ±60kPa | G | 0.05/0.1 |
| 14 | ±100kPa | G | 0.02/0.05 |
| 15 | (-100 ~160) kPa | Generalis Litterarum | 0.02/0.05 |
| 16 | (-100 ~250) kPa | Generalis Litterarum | 0.02/0.05 |
| 17 | (-100 ~400) kPa | Generalis Litterarum | 0.02/0.05 |
| 18 | (-100 ~600) kPa | Generalis Litterarum | 0.02/0.05 |
| 19 | (-0.1~1)Mpa | Generalis Litterarum | 0.02/0.05 |
| 20 | (-0.1~1.6)Mpa | Generalis Litterarum | 0.02/0.05 |
| 21 | (-0.1~2.5)Mpa | Generalis Litterarum | 0.02/0.05 |
Adnotationes:
1. Spatium partiale pressionem omnino facere potest
2. Spatium compensationis temperaturae automaticae: (-20~50℃)
3. Medium translationis pressionis requirit non corrosivum